FOREWORD
BI0 cabinets provide vertical laminar-flow. Most of the air is recirculated through the HEPA filter, with part being expelled into the environment through the exhaust outlet. BI0 cabinets operate under negative pressure after sterilisation through the HEPA filter and provide Class 100 air according to Federal Standard 209e. They therefore protect the product from micro-organisms and air borne contaminants. The balanced, perforated work surface and design of the working aperture provide an excellent flow laminarity at the work level.
APPLICATIONS
These cabinets have been specifically designed to meet the demands of modern industrial laboratories, primarily for handling cell cultures and no-hazardous biological agents, when it is necessary to ensure product protection.
BI0 cabinets have applications in a wide range of disciplines, including:
- Quality control of pharmaceuticals and food industries
- Microbiology
- Cell culture
- Sterile handling
SPECIFICATIONS
Construction for mod. BI0 48/60/72: AISI-304 stainless steel to provide hygienic conditions.
Construction for mod. BI0 48M/60-M/72-M: corrosion resistant epoxy powder painted steel.
Work surface: glazed AISI-stainless steel completely perforated to ensure flow laminarity and air recycling at the work surface level.
Transparent side walls: safety glass.
Absolute filter: HEPA filter, tested with D. O. P. aerosol. Their efficiency, higher than 99.999% at 0.3 micron particle size -0.001%penetration -ensures performance exceeding requirements of U. S. Federal Standard 209 e, British Standard 5295, VDI-2083 and AS 1386-1976.
Motor blower: direct coupled motor, electronic speed controlled to maintain a constant air flow of 0.45 m/sec, and compensate for a partially clogged filter up to a maximum plenum pressure of 40 mm of water.
Power Supply: 220 V, 50 Hz, single phase. Other voltages available on request.
Lighting: remote mounted fluorescent light, outside sterile area.
Front window: safety glass window with camlock.
Spillage tray: a painted spillage tray for mod. BIO-M or AISI-304 stainless steel spillage tray for mod. BI0 beneath the work surface collects spilt liquid.
Additional Features: the work area is equipped with a manual gas tap and IP-44 electrical socket. |
| |
|
Code |
Model |
Useful
Dim. mm |
Overall
Dim.
mm |
Power |
Weight |
| |
|
W |
H |
D |
W |
H |
D |
Kw |
Kg |
F72700350 |
BIO 48 |
1220 |
640 |
580 |
1312 |
1430 |
765 |
0.6 |
150 |
F72700400 |
BIO 48-M |
1220 |
640 |
580 |
1312 |
1430 |
765 |
0.6 |
150 |
F72700370 |
BIO 60 |
1525 |
640 |
580 |
1617 |
1430 |
765 |
0.7 |
180 |
F72700410 |
BIO 60-M |
1525 |
640 |
580 |
1617 |
1430 |
765 |
0.7 |
180 |
F72700360 |
BIO 72 |
1830 |
640 |
580 |
1922 |
1430 |
765 |
0.8 |
200 |
F72700420 |
BIO 72-M |
1830 |
640 |
580 |
1922 |
1430 |
765 |
0.8 |
200 |
|
Model |
Exhausted Air
m3/hr |
Temperature rise |
Noise level
dBA |
Lighting
lux |
Vibrations
mm RMS |
|
BIO-48 |
150 |
< 3°C |
<59 |
>900 |
<0.006 |
|
BIO-60 |
160 |
< 3°C |
<59 |
>900 |
<0.006 |
|
BIO-72 |
300 |
< 3°C |
<59 |
>900 |
<0.006 |
|
|
| |